FTIR vs. Mass Spectrometry: How Peptide Identity Is Confirmed
Purity tells you how much of a sample is the main compound. Identity testing answers an earlier question: is the main compound actually the peptide on the label? A sample can be 99% pure and still be the wrong molecule. That is why a complete certificate of analysis (COA) pairs an HPLC purity result with a separate identity test. The two most common identity methods for research peptides are Fourier-transform infrared spectroscopy (FTIR) and mass spectrometry (MS).
What FTIR measures
FTIR shines infrared light through or onto a sample and records which wavelengths the sample absorbs. Chemical bonds absorb infrared light at characteristic frequencies, so the resulting spectrum is a pattern of absorption bands that reflects the molecule’s functional groups. For peptides, the most prominent features are the amide bands from the peptide backbone, along with bands from side chains and any modifications.
Identity is confirmed by comparing the sample’s spectrum with a reference spectrum of the same compound. If the bands line up in position and relative intensity, the result is reported as a match, often with a correlation or similarity score.
- Strengths: fast, inexpensive, needs little sample, and gives a recognisable fingerprint of the material as a whole.
- Limits: it depends on having a good reference spectrum, and peptides with similar sequences can produce similar spectra. Excipients and moisture also contribute bands, so the lab must account for them.
What mass spectrometry measures
Mass spectrometry ionises the sample and measures the mass-to-charge ratio of the resulting ions. For a peptide, the key result is the observed molecular mass, which is compared with the theoretical mass calculated from its sequence and any modifications. Peptides often appear as several charge states, which the lab deconvolutes to a single molecular mass.
- Strengths: highly specific. A mass that matches the theoretical value to within a small tolerance is strong evidence of the correct sequence and modifications. When coupled with liquid chromatography (LC-MS), it can also help identify impurity peaks.
- Limits: more expensive and specialised than FTIR. A correct total mass does not by itself prove the amino acids are in the right order; sequence confirmation needs fragmentation (MS/MS) data.
Reading identity results on a COA
| What to look for | Why it matters |
|---|---|
| The method named (FTIR, MS, LC-MS) | Tells you which kind of evidence the identity claim rests on. |
| A clear pass/fail or “conforms” result against a specification | Identity should be judged against a defined reference, not stated without one. |
| The spectrum itself | A COA that includes the FTIR spectrum or mass spectrum lets you see the data behind the conclusion. |
| Observed vs. theoretical mass (for MS) | The two values should agree closely, within the lab’s stated tolerance. |
| Lot number | The identity result applies only to the lot tested. |
Identity and purity work together
HPLC purity and identity testing answer different questions, and neither replaces the other. HPLC shows that one component dominates the sample; FTIR or MS shows that the dominant component is the expected peptide. A COA that reports both, for the same lot, gives a far more complete picture than either result alone.
How we handle identity testing
Every peptide lot Proven Peptides sells is tested by an independent third-party laboratory for HPLC purity and quantity and for FTIR identity. The full report for each lot, including the spectra, is published on our Lab Results page.